#PAGE_PARAMS# #ADS_HEAD_SCRIPTS# #MICRODATA#

Spectral characteristics of urine from patients with end-stage kidney disease analyzed using Raman Chemometric Urinalysis (Rametrix)


Authors: Ryan S. Senger aff001;  Meaghan Sullivan aff001;  Austin Gouldin aff001;  Stephanie Lundgren aff001;  Kristen Merrifield aff001;  Caitlin Steen aff001;  Emily Baker aff001;  Tommy Vu aff002;  Ben Agnor aff001;  Gabrielle Martinez aff001;  Hana Coogan aff001;  William Carswell aff001;  Varun Kavuru aff004;  Lampros Karageorge aff004;  Devasmita Dev aff004;  Pang Du aff005;  Allan Sklar aff006;  James Pirkle, Jr aff007;  Susan Guelich aff008;  Giuseppe Orlando aff009;  John L. Robertson aff003
Authors place of work: Department of Biological Systems Engineering, Virginia Tech, Blacksburg, Virginia, United States of America aff001;  Department of Chemical Engineering, Virginia Tech, Blacksburg, Virginia, United States of America aff002;  DialySenors, Inc., Blacksburg, Virginia, United States of America aff003;  Veteran Affairs Medical Center, Salem, Virginia, United States of America aff004;  Department of Statistics, Virginia Tech, Blacksburg, Virginia, United States of America aff005;  Lewis-Gale Medical Center, Salem, Virginia, United States of America aff006;  Department of Internal Medicine–Nephrology, Wake Forest University Baptist Medical Center, Winston-Salem, North Carolina, United States of America aff007;  Valley Nephrology Associates, Roanoke, Virginia, United States of America aff008;  Department of Surgical Sciences–Transplant, Wake Forest University Baptist Medical Center, Winston-Salem, North Carolina, United States of America aff009;  Department of Biomedical Engineering and Mechanics, Virginia Tech, Blacksburg, Virginia, United States of America aff010;  Virginia Tech-Carilion School of Medicine and Research Institute, Blacksburg, Virginia, United States of America aff011
Published in the journal: PLoS ONE 15(1)
Category: Research Article
doi: https://doi.org/10.1371/journal.pone.0227281

Summary

Raman Chemometric Urinalysis (RametrixTM) was used to discern differences in Raman spectra from (i) 362 urine specimens from patients receiving peritoneal dialysis (PD) therapy for end-stage kidney disease (ESKD), (ii) 395 spent dialysate specimens from those PD therapies, and (iii) 235 urine specimens from healthy human volunteers. RametrixTM analysis includes spectral processing (e.g., truncation, baselining, and vector normalization); principal component analysis (PCA); statistical analyses (ANOVA and pairwise comparisons); discriminant analysis of principal components (DAPC); and testing DAPC models using a leave-one-out build/test validation procedure. Results showed distinct and statistically significant differences between the three types of specimens mentioned above. Further, when introducing “unknown” specimens, RametrixTM was able to identify the type of specimen (as PD patient urine or spent dialysate) with better than 98% accuracy, sensitivity, and specificity. RametrixTM was able to identify “unknown” urine specimens as from PD patients or healthy human volunteers with better than 96% accuracy (with better than 97% sensitivity and 94% specificity). This demonstrates that an entire Raman spectrum of a urine or spent dialysate specimen can be used to determine its identity or the presence of ESKD by the donor.

Keywords:

biomarkers – Statistical data – Specimen preparation and treatment – chronic kidney disease – Medical dialysis – Urine – Metabolomics – Principal component analysis

Introduction

The chemical composition, physical characteristics, and types/amounts of suspended materials in urine change when kidney (and systemic) disease is present [14]. In this study, Raman Chemometric Urinalysis (RametrixTM) [58] was used to determine if differences in molecular spectra could be detected in the following specimen types: (i) urine from healthy human volunteers, (ii) urine from patients undergoing peritoneal dialysis (PD) therapy for end-stage kidney disease (ESKD) and (iii) spent dialysate from these patients receiving PD therapy. RametrixTM relies on Raman spectroscopy and a biological region of the spectrum (400–1,800 cm-1) that is composed of spectral signatures of the thousands of molecules known to the urine metabolome [9,10]. Furthermore, RametrixTM uses off-the-shelf Raman spectrometers, which are becoming more affordable, low-profile, and conducive to clinical laboratory use [11].

In the past, sophisticated analyses, including mass spectrometry, liquid/gas chromatography, and kinetic nephelometry have been used to detect analytes (i.e., “biomarkers”) in urine associated with metabolism or disease [3,1215]. In addition, urine contains byproducts of therapeutics that help clinicians monitor and adjust therapies [16,17] and can indicate environmental and occupational toxin exposure [18].

Urine biomarker studies to detect chronic kidney disease have been conducted almost exclusively in research settings. For example, Zürbig and co-workers [19] used capillary electrophoresis, coupled online to electrospray ionization time-of-flight mass spectrometry (CE-MS), to identify polypeptide patterns in human urine. However, proteomic spectral patterns in urine, as biomarkers for kidney disease, have not been accepted widely as diagnostic tools, and proteomic patterns identified in that study could not be translated to specific molecules with physiological and pathophysiological significance. This significantly dampened enthusiasm for proteomic spectral pattern recognition as a tool to diagnose and study genitourinary tract disease.

For these and other reasons, biomarker and “-omics” technologies are used rarely (or are not readily available) by caregivers in patient care settings. This is due to expense, the daunting requirement for advanced technology (such as mass spectrometry), expertise required for interpretation of results, and a lack of assay validation with large datasets of normal and abnormal specimens. In fact, the complexity of both acute and chronic kidney diseases, and other genitourinary tract pathologies, makes large dataset sampling and validation both unlikely and cost-prohibitive [2023]. RametrixTM, on the other hand, provides robust spectral information about the urine metabolome, using instrumentation that is inexpensive. To achieve useful results, minimal technician training is required, and RametrixTM software automates analysis. This has enabled studies, such as this and others [58,2427], using very large datasets, to discover spectral differences between normal and abnormal (e.g., diseased, stressed, decayed, chemically treated, etc.) samples.

Previously, we reported the Rametrix analysis of 235 urine specimens obtained from consented, healthy, human volunteers [8]. This work identified common Raman spectral characteristics seen in urine specimens obtained from healthy persons of both sexes and between 18–70 years of age. We focused on determining the variation (or lack thereof) in spectral signatures from several subsets of individuals over 30 days, comparing data from single time-point (first voided sample in the morning) collections. As part of this work, we noted significant effects of sex and age of the donor but negligible effects of menstruation on Raman spectral characteristics.

Here, we describe the results of Rametrix analysis of 362 urine specimens collected from patients with ESKD, who had residual renal function and were undergoing PD treatments. We also describe RametrixTM analysis of 395 specimens of spent dialysate from those patients undergoing PD treatments. These were then compared with the 235 urine specimens from healthy human volunteers [8] using RametrixTM. All collections and data analysis were devoted to testing the hypothesis that “Rametrix analysis of urine specimens can discern significant molecular composition differences between (i) PD patient urine and spent dialysate and (ii) PD patient urine and urine from healthy individuals.” The study to discern the molecular differences between PD patient urine and spent dialysate was designed to demonstrate the capabilities of RametrixTM analysis and provide a basis for future studies that look at patient-specific molecules that are removed by PD or the kidneys. The study to discern PD patient urine from urine of healthy individuals was designed to identify new patients who should receive PD therapy and to be able to monitor the progress of those patients receiving PD therapy.

Materials and methods

Informed consent

Informed written consent for the collection of urine specimens from healthy human volunteers was obtained under research protocol VT15-703, approved and administered by the Virginia Tech Institutional Review Board. Informed written consent for the collection of urine and dialysate specimens from patients undergoing dialysis therapies was obtained under research protocol RPP/177151.2, approved and administered by Frenova (Fresenius Renal Research; 920 Winter Street, Waltham, MA 02451). In accordance with these protocols, specimens were de-identified and assigned a code at the time of collection.

Description of study population and sampling

Three groups of specimens were compared in this study: (i) urine from patients undergoing PD therapy for ESKD, (ii) the spent dialysate from those PD therapies, and (iii) urine from healthy human volunteers. A full analysis of the healthy human volunteer urine dataset has been published [8]. Briefly, 235 urine specimens were collected from 48 (39 females, 9 males) healthy human volunteers with no history or evidence of renal disease. Volunteers were also free of infectious or degenerative disease at the time of sample collection. The age range of the healthy volunteer population was 18–70 years; 87.5% of volunteers were of ages 19–22 years, and the median age was 21 years.

For patients undergoing PD therapies, 362 urine specimens were collected from 96 patients, and 395 spent dialysate specimens from 115 patients comprised the dataset. Patients had advanced ESKD and were undergoing PD treatment. Patients ranged in age from 24–90 years old. The mean age was 60 years, and the median age was 63.5 years. Multiple collections (4–8 separate collections) were available from multiple patients, allowing repetitive measurements and correlations over a protracted course of PD therapy (18 months).

Specimen collection and storage

Specimens were collected at the time of routine PD adequacy evaluation (generally every 1–3 months) over a period ranging from 18–24 months. For routine adequacy testing, patients collected all urine produced in a 24-hour period and also collected all spent dialysate from multiple cycles of treatment that occurred over 24 hours. These urine samples and spent dialysate collections were brought to the dialysis center, where aliquots of urine and spent dialysate were transferred into sterile specimen cups and then immediately frozen to -15°C. Both urine and spent dialysate specimens were stored at this temperature until analyzed. Urine specimens from healthy human volunteers were stored immediately at -35°C until analyzed.

We previously determined the suitability of collection and storage conditions in a separate study of urine stability [7] and adhered to the guidelines set forth in that study. Unused portions of urine and spent dialysate specimens were stored at -35°C for the duration of the study and re-analyzed, as needed.

Analytical standards

Surine Urine Negative Control (Dyna-Tek Industries, Lenexa, KS) was used as a control standard for urinalysis. Unused dialysate (obtained from Valley Nephrology Associates; Roanoke, VA) was also used as a reference control in this study.

Raman methodology and measurements

Previously published experimental methods were used [6, 7]. Briefly, an Agiltron PeakSeeker dispersive Raman spectrometer (Woburn, MA) was used, and all specimens were Raman scanned as bulk liquid samples in 1.5 mL glass vials at 25°C using 785 nm (30 mW) laser excitation for 30 s with spectral resolution of 8 cm-1. A minimum of 10 scans were collected per specimen and averaged.

Computational methodology

Previously published computational methods were also used [7,8]. Spectral processing and analyses were performed with the RametrixTM LITE [5], RametrixTM PRO [6], and Statistics and Machine Learning Toolboxes were used with MATLAB r2018A (The MathWorks, Inc.; Natick, MA). Raman spectra were truncated to 400–1,800 cm-1, baseline corrected using the Goldindec algorithm [28], and vector normalized. Principal component analysis (PCA) and discriminant analysis of principal components (DAPC) models were constructed using the RametrixTM LITE Toolbox, and DAPC models were tested by leave-one-out analysis with the RametrixTM PRO Toolbox. 1-Way ANOVA and pairwise comparisons using Tukey’s honestly significant difference (HSD) procedure were performed in MATLAB.

Public availability

The RametrixTM LITE Toolbox is available through GitHub under an MIT licensing agreement (https://github.com/SengerLab/RametrixLITEToolbox). The RametrixTM PRO Toolbox is also available through GitHub under similar licensing agreement (https://github.com/SengerLab/RametrixPROToolbox). Raman scans of urine from ESKD patients are available with these tools. Data used for statistical analyses are also available through GitHub (https://github.com/SengerLab/Raman-Scans/tree/ESKD).

Results

Relevant questions

Raman spectroscopy and RametrixTM analysis was used to answer the following questions:

  • Are PD patient urine and spent dialysate different?

  • Are PD patient urine and urine from healthy volunteers different?

  • Can RametrixTM identify “unknown” specimens correctly?

Raman spectroscopy of PD patient urine and spent dialysate

Raman spectra from 362 urine specimens and 395 spent dialysate specimens from PD patients were averaged (per specimen), baseline corrected using the Goldindec algorithm [27], and vector normalized. These were plotted together in Fig 1A (PD patient urine) and Fig 1B (spent dialysate). Raman spectra of the 235 urine specimens from healthy human volunteers were processed similarly and published elsewhere [8]. Apparent in Fig 1, there are observable differences between Raman spectra of PD patient urine and spent dialysate. In particular, there are clear differences in the urea (1,002 cm-1) [10,29] content of these specimens. While all specimens seemed to show a basic spectral signature of urine or spent dialysate, there were considerable differences between the individual spectra within these specimen types. These were explored further by PCA and statistical tests to provide quantitative metrics.

Raman spectra of PD patient urine and spent dialysate.
Fig. 1. Raman spectra of PD patient urine and spent dialysate.
(A) Averaged, baselined, and vector normalized Raman spectra from 362 urine specimens obtained from patients receiving PD therapy for ESKD. (B) Averaged, baselined, and vector normalized Raman spectra from 395 spent PD dialysate specimens.

Principal component analysis

The PD patient urine specimens were compared against the spent dialysate specimens using PCA in the RametrixTM LITE Toolbox. The first two principal components are plotted in Fig 2A. Here, significant separation is observed between the two specimen types (i.e., PD urine and spent dialysate). PCA was applied with two controls: (i) SurineTM as a urinalysis standard and (ii) unused dialysate. SurineTM clustered with the urine specimens, and the unused dialysate clustered with the spent dialysate, showing some similarity between these specimen types. The RametrixTM LITE Toolbox also identifies Raman shifts that lead to the separation of clusters in PCA [5]. These can be traced back to individual molecules by scanning individual standards, metabolomic knowledge [9], and spectral libraries [10]. For the PD patient urine and spent dialysate dataset, these Raman shift contributions are shown in Fig 2B. Contributions from the top four principal components are shown, and together, these represent over 92% of the dataset variance. Again, the Raman shift at 1,002 cm-1 was the most dominant, present in all principal components, and is representative of urea in urine and dialysate specimens. Other notable Raman shifts in Fig 2B include creatinine (680 cm-1) [30,31] and glucose (1,071 cm-1; 1,117 cm-1; others) [8,32]. Research is ongoing to validate more Raman shifts in this and similar plots using Raman scans of standards and metabolomic analysis. We note, however, the chemometric approach of RametrixTM (described in the following sections) allow meaningful results to be obtained without the assignment of analytes to individual Raman bands.

PCA of PD patient urine and spent dialysate.
Fig. 2. PCA of PD patient urine and spent dialysate.
(A) PCA results for Raman spectra of 362 urine specimens obtained from patients receiving PD therapy for ESKD and 395 spent dialysate specimens. (B) Contributions of Raman shifts leading to separations among principal components.

The PD patient urine specimens were also compared against the urine specimens from healthy human volunteers by PCA. These results are shown in Fig 3A and show the separation of clusters between the two specimen types, suggesting significant spectral differences and molecular compositions. This time, the SurineTM urinalysis control clustered with urine specimens from healthy volunteers, rather than those from PD patients. The spectral differences (i.e., signal intensities at each Raman shift) that lead to the observed separation between urine specimens from PD patients and healthy volunteers in PCA are given in Fig 3B. The top four principal components represent more than 94% of the dataset variance in this case. Here, urea (1,002 cm-1) is dominant in the first principal component (PC 1). Again, creatinine and glucose are apparent in Fig 3B, and additional Raman shift contributions are present relative to the comparison of PD patient urine and spent dialysate (Fig 2B).

PCA of PD patient urine and urine from healthy individuals.
Fig. 3. PCA of PD patient urine and urine from healthy individuals.
(A) PCA results for Raman spectra of 362 urine specimens obtained from patients receiving PD therapy for ESKD and 235 urine specimens from healthy individuals. (B) Contributions of Raman shifts leading to separations among principal components.

Statistical analyses

The entire dataset (consisting of Raman scans of healthy human volunteer urine, PD patients urine, and spent dialysate) was analyzed by 1-way ANOVA to determine if the type of specimen was statistically significant. To do this, the spectra were each quantified to a single numerical value through calculation of the total principal component distance (TPD). This calculation has been explained and demonstrated previously [7,8]. Briefly, TPD represents how closely the Raman spectrum of a specimen resembles that of SurineTM. To calculate this, the distance formula is applied between the top four principal components of a specimen and those of SurineTM. This procedure reduces a data-rich Raman spectrum down to a single numerical value, which allows statistical tests to be applied. In particular, a 1-way ANOVA test of TPD values for all specimen types (i.e., healthy human urine, PD patient urine, or spent dialysate) returned a p-value less than 0.001, which confirmed that the type of specimen was statistically significant. Results of pairwise comparison tests with Tukey’s HSD procedure are shown in Table 1. Statistical significance (p < 0.001) was obtained when comparing (i) PD patient urine with spent dialysate, (ii) PD patient urine with healthy human volunteer urine, and (iii) PD patient spent dialysate with healthy human volunteer urine. Results confirm these types of specimens are all different from one another. Other pairwise comparisons did not return statistical significance likely because only single controls (SurineTM and unused dialysate) were included in the dataset.

Tab. 1. Pairwise comparisons using Tukey’s HSD procedure.
Pairwise comparisons using Tukey’s HSD procedure.

Discriminant analysis of principal components models

The RametrixTM LITE Toolbox was used to generate DAPC models for datasets consisting of Raman scans of: (i) PD patient urine and spent dialysate and (ii) PD patient urine and urine from healthy human volunteers. SurineTM was included as a control in both models, and unused dialysate was used as an additional control for the model containing PD patient spent dialysate. The RametrixTM PRO Toolbox was then used to evaluate the predictive capabilities of these models using a leave-one-out build/test validation routine. DAPC model clustering is shown in Fig 4 for both models when 50 principal components were used in model construction. Good separation of clusters (with some overlap) was observed in both cases. SurineTM was separated from all clusters in both models, and the unused dialysate standard clustered with the PD patient spent dialysate specimens.

DAPC of PD patient urine, spent dialysate, and urine from healthy individuals.
Fig. 4. DAPC of PD patient urine, spent dialysate, and urine from healthy individuals.
DAPC results for models made with 50 principal components. (A) 362 urine specimens obtained from patients receiving PD therapy for ESKD and 395 spent dialysate specimens. (B) 362 urine specimens obtained from patients receiving PD therapy for ESKD and 235 urine specimens from healthy individuals.

The prediction capabilities (from “leave-one-out” build/test routines) of the DAPC models are shown in Tables 2 and 3. The DAPC models were built with different numbers of principal components to ensure enough dataset variance was included in the models and to test for model overfitting. DAPC models were evaluated in terms of prediction accuracy (the percentage of data points predicted correctly), sensitivity (the true positive percentage), and specificity (the true negative percentage). Results in Table 2 convey RametrixTM can determine the identity of an “unknown” specimen from a PD patient as being either urine or spent dialysate with very high confidence. Greater than 98% accuracy, sensitivity, and specificity were obtained for a DAPC model consisting of 10 principal components. For all DAPC models tested, the accuracy, sensitivity, and specificity values exceeded 97%. This high level of confidence in identifying the type of sample (i.e., urine or spent dialysate) is unsurprising given the clear differences in Raman spectra shown in Fig 1. For determining whether a urine specimen originated from a PD patient or healthy human volunteer, RametrixTM prediction results are given in Table 3. Better than 96% prediction accuracy (with better than 97% sensitivity and 94% specificity) was obtained for the DAPC model constructed with 11 principal components. Using 50 principal components led to increased accuracy and sensitivity with decreased specificity, which is characteristic of model over-fitting.

Tab. 2. RametrixTM PRO results showing the ability to predict whether an unknown specimen from a PD patient is urine or spent dialysate.
Rametrix<sup>TM</sup> PRO results showing the ability to predict whether an unknown specimen from a PD patient is urine or spent dialysate.
Tab. 3. RametrixTM PRO results showing the ability to predict whether an unknown urine specimen came from a PD patient or healthy human volunteer.
Rametrix<sup>TM</sup> PRO results showing the ability to predict whether an unknown urine specimen came from a PD patient or healthy human volunteer.

Discussion

RametrixTM has demonstrated the ability to discern effectively among (i) urine from PD patients (ESKD), (ii) spent dialysate from their PD therapies, and (iii) urine from healthy human volunteers. Cluster separations (according to specimen type) were readily apparent in PCA and DAPC model plots, and the conversion of spectral data to TPD values for statistical analyses also confirmed these differences were statistically significant. We have begun the process of identifying molecules responsible for these differences, and we hypothesize this may result in a new set of biomarkers for ESKD and earlier stages of chronic kidney disease. However, we were able to show that the entire Raman spectrum of a specimen can be used (i.e., chemometrics) to determine its type (i.e., urine or dialysate) or the state of the donor (i.e., healthy human or PD patient). The leave-one-out build/test validations of Tables 2 and 3 are particularly important because they describe how well RametrixTM can perform with “unknown” specimens. Of course, the long-term vision with RametrixTM is not to be able to discern whether an unknown specimen is urine or dialysate but to be able to screen for the presence of incipient disease and patient-specific PD responses. For example, RametrixTM could be used to determine whether PD therapies are patient-specific and if there are variations among successive treatments. If so, how do these affect long-term patient outcomes? Comparing urine from PD patients to that of healthy human volunteers is a first crucial step in this process. The PD patients used in this study have ESKD; thus, the differences between urine from these patients and healthy human volunteers should be significant and easily detectable by RametrixTM, as was confirmed by this study. The next steps in RametrixTM development are to be able to detect earlier stages (i.e., G1-4) and track patient progress over longer periods of time.


Zdroje

1. Echeverry G, Hortin GL, Rai AJ. Introduction to urinalysis: historical perspectives and clinical application. Methods Mol Biol Clifton NJ. 2010;641: 1–12. doi: 10.1007/978-1-60761-711-2_1 20407938

2. Laposata M, editor. Laboratory Medicine: The diagnosis of disease in the clinical laboratory. 2nd ed. New York: McGraw-Hill Education; 2014.

3. Levey AS, Becker C, Inker LA. Glomerular filtration rate and albuminuria for detection and staging of acute and chronic kidney disease in adults: a systematic review. JAMA. 2015;313: 837–846. doi: 10.1001/jama.2015.0602 25710660

4. Mundt L, Shanahan K. Graff’s textbook of urinalysis and body fluids. 3rd ed. Philadelphia, PA: Wolters-Klower; 2016.

5. Fisher AK, Carswell WF, Athamneh AIM, Sullivan MC, Robertson JL, Bevan DR, et al. The RametrixTM LITE Toolbox v1.0 for MATLAB®. J Raman Spectrosc. 2018;49: 885–896. doi: 10.1002/jrs.5348

6. Senger R, Robertson J. The RametrixTM PRO Toolbox v1.0 for MATLAB®. In Press. 2019

7. Senger R, Du P, DeLaTorre Campos D, Carswell W, Webster K, Sullivan M, et al. Assessment of urine specimen storage conditions using Raman Chemometric UrinalysisTM (RametrixTM). Under Review. 2019

8. Senger R, Kavuru V, Sullivan M, Gouldin A, Lundgren S, Merrifield K, et al. Spectral characteristics of urine specimens from healthy human volunteers analyzed using Raman chemometric urinalysis (Rametrix). PLOS ONE 2019; doi: 10.1371/journal.pone.0222115 31560690

9. Bouatra S, Aziat F, Mandal R, Guo AC, Wilson MR, Knox C, et al. The Human Urine Metabolome. PLOS ONE. 2013;8: e73076. doi: 10.1371/journal.pone.0073076 24023812

10. Movasaghi Z, Rehman S, Rehman DIU. Raman Spectroscopy of Biological Tissues. Appl Spectrosc Rev. 2007;42: 493–541. doi: 10.1080/05704920701551530

11. Pence I, Mahadevan-Jansen A. Clinical instrumentation and applications of Raman spectroscopy. Chem Soc Rev. 2016;45: 1958–1979. doi: 10.1039/c5cs00581g 26999370

12. Gabler J, Wang S. Quantification of metanephrine and normetanephrine in urine using liquid chromatography-tandem mass spectrometry. Methods Mol Biol Clifton NJ. 2016;1378: 149–157. doi: 10.1007/978-1-4939-3182-8_17

13. Lenders JWM, Williams TA, Reincke M, Gomez-Sanchez CE. DIAGNOSIS OF ENDOCRINE DISEASE: 18-Oxocortisol and 18-hydroxycortisol: is there clinical utility of these steroids? Eur J Endocrinol. 2018;178: R1–R9. doi: 10.1530/EJE-17-0563 28904009

14. Willrich MAV, Katzmann JA. Laboratory testing requirements for diagnosis and follow-up of multiple myeloma and related plasma cell dyscrasias. Clin Chem Lab Med. 2016;54: 907–919. doi: 10.1515/cclm-2015-0580 26509779

15. Bax C, Taverna G, Eusebio L, Sironi S, Grizzi F, Guazzoni G, et al. Innovative diagnostic methods for early prostate cancer detection through urine analysis: A review. Cancers. 2018;10. doi: 10.3390/cancers10040123 29670060

16. Alves RC, Fernandes RP, Eloy JO, Salgado HRN, Chorilli M. Characteristics, Properties and analytical methods of Paclitaxel: A review. Crit Rev Anal Chem. 2018;48: 110–118. doi: 10.1080/10408347.2017.1416283 29239659

17. Mahajan G. Role of Urine Drug Testing in the Current Opioid Epidemic. Anesth Analg. 2017;125: 2094–2104. doi: 10.1213/ANE.0000000000002565 29189366

18. Richter P, Faroon O, Pappas RS. Cadmium and cadmium/zinc ratios and Tobacco-Related morbidities. Int J Environ Res Public Health. 2017;14. doi: 10.3390/ijerph14101154 28961214

19. Zürbig P, Mischak H. Capillary electrophoresis coupled to mass spectrometry for biomarker discovery and diagnosis of kidney diseases. Contrib Nephrol. 2008;160: 107–126. doi: 10.1159/000125938 18401165

20. Reichel RR. Acute kidney injury: quoi de neuf? Ochsner J. 2014;14: 359–368. 25249802

21. Prakoura N, Chatziantoniou C. Periostin in kidney diseases. Cell Mol Life Sci CMLS. 2017;74: 4315–4320. doi: 10.1007/s00018-017-2650-6 28884334

22. Shardlow A, McIntyre NJ, Fraser SDS, Roderick P, Raftery J, Fluck RJ, et al. The clinical utility and cost impact of cystatin C measurement in the diagnosis and management of chronic kidney disease: A primary care cohort study. PLoS Med. 2017;14: e1002400. doi: 10.1371/journal.pmed.1002400 29016597

23. Edelstein C. Biomarkers of acute kidney injury. Biomarkers of kidney disease. 2nd ed. Academic Press; 2017.

24. Athamneh AIM, Alajlouni RA, Wallace RS, Seleem MN, Senger RS. Phenotypic Profiling of Antibiotic Response Signatures in Escherichia coli Using Raman Spectroscopy. Antimicrob Agents Chemother. 2014;58: 1302–1314. doi: 10.1128/AAC.02098-13 24295982

25. Freedman BG, Zu TNK, Wallace RS, Senger RS. Raman spectroscopy detects phenotypic differences among Escherichia coli enriched for 1-butanol tolerance using a metagenomic DNA library. Biotechnol J. 2016;11: 877–889. doi: 10.1002/biot.201500144 26814030

26. Zu TNK, Athamneh AIM, Collakova E, Robertson J, Hawken T, Aardema C, et al. Assessment of ex vivo perfused liver health by Raman spectroscopy. J Raman Spectrosc. 2015;46: 551–558. doi: 10.1002/jrs.4688

27. Zu TNK, Athamneh AIM, Wallace RS, Collakova E, Senger RS. Near-Real-Time Analysis of the Phenotypic Responses of Escherichia coli to 1-Butanol Exposure Using Raman Spectroscopy. J Bacteriol. 2014;196: 3983–3991. doi: 10.1128/JB.01590-14 25157078

28. Liu J, Sun J, Huang X, Li G, Liu B. Goldindec: A novel algorithm for Raman spectrum baseline correction. Appl Spectrosc. 2015;69: 834–842. doi: 10.1366/14-07798 26037638

29. Premasiri WR, Clarke RH, Womble ME. Urine analysis by laser Raman spectroscopy. Lasers Surg Med. 2001;28: 330–334. doi: 10.1002/lsm.1058 11344513

30. Frank CJ, McCreery RL, Redd DCB. Raman Spectroscopy of normal and diseased human breast tissues. Anal Chem. 1995;67: 777–783. doi: 10.1021/ac00101a001 7762814

31. Gniadecka M, Wulf HC, Mortensen NN, Nielsen OF, Christensen DH. Diagnosis of Basal cell carcinoma by Raman spectroscopy. J Raman Spectrosc. 1997;28: 125–129. doi: 10.1002/(SICI)1097-4555(199702)28:2/3<125::AID-JRS65>3.0.CO;2-#

32. Krafft C, Neudert L, Simat T, Salzer R. Near infrared Raman spectra of human brain lipids. Spectrochim Acta A Mol Biomol Spectrosc. 2005;61: 1529–1535. doi: 10.1016/j.saa.2004.11.017 15820887


Článok vyšiel v časopise

PLOS One


2020 Číslo 1
Najčítanejšie tento týždeň
Najčítanejšie v tomto čísle
Kurzy

Zvýšte si kvalifikáciu online z pohodlia domova

Získaná hemofilie - Povědomí o nemoci a její diagnostika
nový kurz

Eozinofilní granulomatóza s polyangiitidou
Autori: doc. MUDr. Martina Doubková, Ph.D.

Všetky kurzy
Prihlásenie
Zabudnuté heslo

Zadajte e-mailovú adresu, s ktorou ste vytvárali účet. Budú Vám na ňu zasielané informácie k nastaveniu nového hesla.

Prihlásenie

Nemáte účet?  Registrujte sa

#ADS_BOTTOM_SCRIPTS#