A method of compliance measurement and gastight testing in models of the respiratory system
Authors: Karel Roubík; Monika Stránská
Authors place of work: Czech Technical University in Prague, Faculty of Biomedical Engineering, Kladno, Czech Republic
Published in the journal: Lékař a technika - Clinician and Technology No. 1, 2014, 44, 9-18
Category: Původní práce
Summary
The article deals with design and verification of an easy-to-use method for compliance measurement in models of the respiratory system. A new term “effective compliance” has been proposed as an additional parameter to already existing adiabatic and isothermal compliances. A method of gastight testing is described as well. All the methods were tested using a set of compensated and uncompensated rigid models of the respiratory system as well as using passive and active lung models with adjustable compliances.
Keywords:
model of the respiratory system, compliance, effective compliance, adiabatic compliance, isothermal compliance
Zdroje
[1] Roubík, K., Ráfl, J.: Poddajnost a resistance v rigidních modelech respirační soustavy. Lékař a technika, 2009, roč. 39, č. 1, s. 32–38. ISSN 0301-5491.
[2] Ráfl, J., Roubík, K.: Fyzické modely respirační soustavy a možnosti jejich využití. Lékař a technika, 2009, roč. 39, č. 1, s. 32–38. ISSN 0301-5491.
[3] IngMar Medical: ASL 5000TM Active Servo Lung, Breathing Simulator, User’s manual, Software 3.2. 2nd Revision, 2007.
[4] Vavřík D., Hájek M., Penhaker M.: Měření a vyhodnocení základních ventilačních režimů a poruch v simulačním ventilačním okruhu. (The measurement and evaluation of basic ventilation modes and failure simulations on ventilation circuit). Lékař a technika, 2008, roč. 38, č. 2, s. 191-194. ISSN 0301-5491.
[5] Penhaker, M., Vavrik, D: The measurement and failure simulations of basic ventilation modes. International Conference on Networking and Information Technology. 2010, Article number 5508451, pp. 554-556.
Štítky
Biomedicína
Článok vyšiel v časopiseLékař a technika
2014 Číslo 1-
Všetky články tohto čísla
- Rational operation of MRI equipment in university hospitals in the Czech Republic
- Individualization of head related transfer function
- Gene expression profiling after angiogenesis inhibitor treatment
- The effect of acetylsalicylic acid on angiogenesis in vitro
- The effect of docetaxel on molecular melting profile of DNA extracted from human breast adenocarcinoma MCF-7 cells
- Metalické nanočástice v prostředí terapeutického ultrazvuku – Studium viability nádorových buněk in vitro
- Metoda měření poddajnosti a těsnosti modelů respirační soustavy pacienta
- Lékař a technika
- Archív čísel
- Aktuálne číslo
- Informácie o časopise
Najčítanejšie v tomto čísle- Metoda měření poddajnosti a těsnosti modelů respirační soustavy pacienta
- Rational operation of MRI equipment in university hospitals in the Czech Republic
- Metalické nanočástice v prostředí terapeutického ultrazvuku – Studium viability nádorových buněk in vitro
- Individualization of head related transfer function
Prihlásenie#ADS_BOTTOM_SCRIPTS#Zabudnuté hesloZadajte e-mailovú adresu, s ktorou ste vytvárali účet. Budú Vám na ňu zasielané informácie k nastaveniu nového hesla.
- Časopisy